distribution switch board

IT system distribution switch board

Energy and protection

  • Safety
  • Changeover device with by-pass switch.

  • Integrated solution
  • Include isolating transformer, changeover devices and isulation monitor.

  • Facilitates access to the technicians

  • Versatile /Flexible
  • Different number of pole circuits brakers.

  • Customize
  • Type of isolating transformers.
    · Three phases or single phase.
    · Power rating 3 to 10kVA.

    DISTRIBUTION SWITCH BOARD

    • Changeover device
    • Isolating transformer
    • Emergency supply unit for surgical lights
    • Insulation monitor
    • Two-pole circuits brakers

    FEATURES OF COMPONENTS

  • Automatic transfer switching device
  • • Blackout <90ms • ModBus connection (optional) • Bypass switch (optional)

  • Isolating transformer
  • • Voltage: 230/400 V~
    • Frequency: 50-60 Hz
    • Safety class: Class I
    • Leak current: < 0,5 mA • Medical use • Power: 3,15 to 10 kVA • Insulating class: F (155º C) • No load current: < 3% • Insulation resistence: 7 MOhmios

  • Emergency supply unit for surgical lights
  • • Switch over system: Instantly
    • Source supply: 100-240 V~/ 50-60Hz
    • Output voltage: 24 to 27,5Vcc
    • Load capacity: 75/150W
    • Emergency supply unit for 2 led light head: 2 hours

  • Two-pole MCBs
  • • At least 8 two-pole MCBs

  • Insulator monitor
  • • Power supply voltage: 100-240 V~ / 50-60 Hz
    • Safety class: Class I
    • Monitoring: Temperature of transformer, Load and History alarms

    IT SYSTEM DISTRIBUTION SWITCH BOARD DETAILS

  • Design
  • Sheet Steel housing.

  • Ambient conditions
  • Conducted waste heat by free convection*
    * Optional: air circularion fan working with temperature sensor.

  • Safety
  • Automatic transfer switching device, with optional bypass switch.

  • Protection
  • Isolating Transformer.

  • Flexible
  • Total adaptation according to the needs of the client and the project.

    TECHNICAL DATA

    Electrical Characteristics
    Input voltage: 400 V 3N~ / 230 V~
    Output voltage: 230 V~ / 230 V~
    Phases: Three phases / Single phase
    Frequency: 50-60 Hz
    Power: 3 to 10 kVA

    Characteristics Board
    Material: Steel
    Surface finish: Epoxy-polyester powder
    IP grade: IP20
    Ik grade: IK06
    Degre of pollution: Grade 2

    Emergency supply unit for surgical lights
    Source supply: 100-240 V~ / 50-60Hz
    Output voltage: 24 to 27,5Vcc
    Load capacity: 75/150W
    Battery: 24Vcc -7AH
    Battery useful life: 5 years

    Ambient Conditions
    Ambient temperature: -10°C to + 60°C
    Relative humidity: 20% to 80%

    Design in according
    IEC 60364-7-710 (Low-voltage electrical installations. Part 7-710: Requirements for special installations or locations. Medical locations), EN 61439-1 (Low-voltage switchgear and controlgear assemblies – Part 1: General rules) and EN 61439-2 (Low-voltage switchgear and controlgear assemblies – Part 2: Power switchgear and controlgear assemblies).

    Electrical Testing in Laboratory
    DEKRA Testing and Certification S.A.U.

    Protection against Electric Shock
    Class I

    CE 0197 – Notify Body
    TüV Rheinland LGA Products GmbH

    Fource Ventilation (optional)

    Fan
    Max flow: 150 / 170 m3/h
    Power consumption: 19 / 22 W

    Temperature sensor
    Precision: +/- 1%
    Range: 0º – 150º